Test Sequencing; In-Depth Sequencing Information; Commands That Sequencing Completes Before The Next Sequence Command - Agilent Technologies 8719D User Manual

Network analyzers
Table of Contents

Advertisement

create a sequence as you are making a measurement. Then when you want to make that same
measurement again, you can recall the sequence and the analyzer will repeat the previous
keystrokes.
The following is a list of some of the key test sequencing features available on the
HP 8719D/20D/22D network analyzer:
Limited decision-making functions increase the versatility of the test sequences you create by
n
allowing you to jump from one sequence to another.
You can create, title, save, and execute up to six sequences.
n
You can save your sequences to a disk using the internal disk drive.
n
You can use the parallel port as a general purpose input/output (GPIO) bus to read five
n
peripheral.
Product note 8753-3 "RF Component Measurements - Applications of the
Note
sequencing. This note was written for the HP 8753B but also applies to the
HP 8719D/20D/22D.

In-Depth Sequencing Information

The analyzer does not allow you to use the following keys in a sequence:

Commands That Sequencing Completes Before the Next Sequence Command

The analyzer completes all operations related to the following commands before continuing
with another sequence command:
number of groups
n
auto scale
n
marker function
n
data + memory
n
recall or save (internal or external)
n
copy list values and operating parameters
n
n
*Wait 0 is the special sequencing function ~~~,~~,~~ with a zero entered for the delay value.