7. Receiver Magnitude Frequency Response
Perform this test to verify how well the analyzer transfers information from the RF to IF,
and how accurately it processes and displays that information.
Analyzer warmup time: 30 minutes.
Specifications
Frequency Range
300 kHz–3 GHz
b
3 GHz–6 GHz
a. At 25 °C ± 5 °C; A, B, R, −10 dBm input.
b. Only for analyzers with Option 006.
Required Equipment
Description
Power Meter
Power Sensor
2-Way power splitter, 50Ω
Attenuator, 10 dB
RF cable set, 50Ω type-N
Adapter, type-N (m) to type-N (m)
Adapter, type-N (f) to type-N (f)
Additional Equipment for Option 006 Analyzers
Power Sensor
Procedure
Input R Magnitude Frequency Response
1. Connect the equipment as shown in
Chapter 2
Magnitude Frequency Response
± 1 dB
± 2 dB
HP/Agilent Part or Model Number
436A/437B/438A or E4418B/4419B
8482A
11667A Option 001
8491A Option 010
11851B
1250-1475
1250-1472
8481A
Figure
2-9.
7. Receiver Magnitude Frequency Response
a
Performance Tests