accepts a calibration factor % for the segment.
brings up the segment modify menu and segment edit
(calibration factor menu) which allows you to enter a power
sensor's calibration factors The calibration factor data entered
in this menu will be stored for power sensor A.
sets the preset state of interpolated error-correction on or off.
brings up the segment modify menu and segment edit
(calibration factor menu) which allows you to enter a power
sensor's calibration factors. The calibration factor data entered
in this menu will be stored for power sensor B.
leads to the select cal kit menu, which is used to select one of
the default calibration kits available for different connector
types. This, in turn, leads to additional menus used to define
calibration standards other than those in the default kits (refer
to Modifying Calibration Kits) When a calibration kit has been
specified, its connector type is displayed in brackets in the
selects the cal kit menu.
selects the HP 85056A/D cal kit.
selects the HP 85056K Cal kit.
selects the 2.92 mm cal kit model.
selects the HP 85033C cal kit.
selects the HP 85033D and HP 85052B/C/D cal kit.
selects the HP 85052C TRL cal kit.
selects the HP 85031B and HP 8505OB/C/D cal kit.
selects the HP 85032B/E and HP 85054B/D cal kit.
selects the HP 85036B/E cal kit.
selects a kit other than those offered by Hewlett-Packard.
this default selection establishes the TRL/LRM LINE/MATCH
standard as the characteristic impedance.
allows you to modify the characteristic impedance of the
system for TRL/LRM calibration.
leads to the calibration menu, which provides several accuracy
enhancement procedures ranging from a simple frequency
response calibration to a full two-port calibration. At the
completion of a calibration procedure, this menu is returned
to the screen, correction is automatically turned on, and the
notation Cor or C2 is displayed at the left of the screen.
is underlined if no calibration has been performed or if the
calibration data has been cleared. Unless a calibration is saved
in memory, the calibration data is lost on instrument preset,
power on, instrument state recall, or if stimulus values are
changed.