PHASE LOCK FAILURE
Error Number The first IF signal was detected at pretune, but phase lock could not be
acquired. Refer to the
7
troubleshooting.
PHASE LOCK LOST
Error Number Phase lock was acquired but then lost. Refer to the
8
Network Anulgzer
PLOT ABORTED
Error Number When you press the LLocal) key, the analyzer aborts the plot in progress.
27
PLOTTER: not on, not connect, wrong addrs
Error Number The plotter does not respond to control. Verify power to the plotter, and check
26
the HP-IB connection between the analyzer and the plotter. Ensure that the
plotter address recognized by the analyzer matches the HP-IB address set on
the plotter itself.
PLOTTER NOT READY-PINCH WHEELS UP
Error Number The plotter pinch wheels clamp the paper in place. If you raise the pinch
wheels, the plotter indicates a "not ready" status on the bus
28
POSSIBLE FALSE LOCK
Error Number Phase lock has been achieved, but the source may be phase locked to the
wrong harmonic of the synthesizer. Perform the source pretune correction
6
routine documented in the "Adjustments and Correction Constants" chapter in
the
HP 8719D/ZODL22D Network Anulgzm semrice Guide.
Error Number The power meter indicates an out-of-range condition. Check the test setup.
116
HP 8719D/ZOD&?ZD Network Aruzlgm Service Guide