Connections for the Second Portion of Conversion Compression Measurement

8. To set the frequency offset mode LO frequency, press:
9. To select the converter type, press:
10. 'Ib select a low-side Lo measurement configuration, press:
In this low-side Lo, up converter measurement, the analyzer source frequency is offset lower
than the receiver frequency. The analyzer source frequency can be determined from the
following equation:
receiver frequency (800 MHz) - Lo frequency (600 MHz) = 200 MHz
The measurement's setup diagram is shown in Figure 3-22.