Figure 3-28. Connections for a Response Calibration
7. Make the connections as shown in Figure 3-29.
Figure 3-29. Connections for a Mixer RF Feedthrough Measurement
8. Connect the external LO source to the mixer's LO port.
9. The measurement results show the mixer's RF feedthrough.
You may see spurious responses on the analyzer trace due to interference
Note
caused by LO to IF leakage in the mixer. This can be reduced with averaging or
by reducing the IF bandwidth.
NETWORK ANALYZER