Solving Problems with Saving or Recalling Files
If you encounter a problem when you are storing llles to disk, or the analyzer internal memory,
check the following list for possible causes:
Look in the analyzer display message area. The analyzer may show a message that will
n
identify the problem. Refer to the "Error Messages" chapter if you view a message.
Make sure that the disk has not been formatted with the LIF-1 (hierarchical file system)
n
extensions as the analyzer does not support this format.
Make sure that you have connected the disk drive to ac power, switched on the power, and
n
an HP-IB cable between the disk drive and the analyzer.
this chapter.
Make sure that the analyzer recognizes the disk (drive) unit that you selected (0 or 1).
n
If the disk drive is an older HP 9122, it may not recognize the newer high density disks.
n
Substitute the HP-IB cable.
n
Substitute the disk drive.
n
Printing, Plotting, and Saving Measurement Results 4 4 3