Response Error-Correction For Transmission Measurements; Standard Connections For Response Error-Correction For Transmission - Agilent Technologies 8719D User Manual

Network analyzers
Table of Contents

Advertisement

1. Press B.
2. Select the type of measurement you want to make.
If you want to make a transmission measurement in the forward direction (&I), press:
q
If you want to make a transmission measurement in the reverse direction (S~Z), press:
q
3. Set any other measurement parameters that you want for the device measurement: power,
number of points, IF bandwidth.
4. To select a response correction, press:
5. Make a "thru" connection between the points where you will connect your device under
test.
Include any adapters or cables that you will have in the device measurement.
That is, connect the standard device where you will connect your device under
test.
Standard Connections for Response Error-Correction for Transmission Measurements
6. To measure the standard, press:
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.
The analyzer underlines the !JZl$%W softkey after it measures the calibration standard, and
computes the error coefficients.
.
NETWORK ANALYZER
Figure 5-2.
. . . : : . . : :
Optimizing Measurement Results 5-l 1