To Use External Calibration - Agilent Technologies 8719D User Manual

Network analyzers
Table of Contents

Advertisement

The m0P B #WW3 sweep mode is the default sweep mode of the analyzer to make two
measurements at the same time. For example, the analyzer can measure A/R and B/R (or Sll
and S21) simultaneously. This mode offers the fastest measurement time.
The other sweep mode is A&~NX@: A MD 's mode, in which one input is measured per
frequency sweep.
You can alternate between sweep modes by going into the (ETJ menu:
For more information, refer to "Alternate and Chop Sweep Modes" in Chapter 6.
Offloading the error correction process to an external PC increases throughput on the
network analyzer. This can be accomplished with remote only commands. Refer to the
HP 8719D/ZOD/ZZD Network Analyzer Prograrnmm's Cuidefor information on how to use
external calibration.
Optimizing Measurement Results