Figure 5-6. Standard Connections for a One Port Reflection Error-Correction
8. To measure the standard, when the displayed trace has settled, press:
If the calibration kit that you selected has a choice between male or female
calibration standards, remember to select the sex that applies to the test port
and not the standard.
The analyzer displays WAIT - @SWUNG CAL STANDARD during the standard measurement.
The analyzer underlines the ~@HJ#&: softkey after it measures the calibration standard.
9. Disconnect the open, and connect a short circuit to the test port.
10. To measure the standard when the displayed trace has settled, press:
The analyzer measures the short circuit and underlines the $&J?I$T softkey.
11. Disconnect the short, and connect an impedance-matched load to the test port. Refer to
"Choosing Calibration Load Standards," located earlier in this chapter.
corresponding to the load used. If a sliding load is used, press ~~~~~~~ to access the
message CAUTION : MORE SLIDES NEEDED will be displayed.
data is measured and the @###@ softkey label is underlined.