Modify The Cal Kit Thru Definition - Agilent Technologies 8719D User Manual

Network analyzers
Table of Contents

Advertisement

With this method it is only necessary to use adapter B. The calibration kit thru de6nition is
modified to compensate for the adapter and then saved as a user kit. However, the electrical
delay of the adapter must first be found.
1. Perform a l-port calibration on PORT 2.
2. Connect adapter B to the test port.
3. Add a short to the open end of the B adapter.
5. Divide the resulting delay measurement by 2.
6. Determine the offset delay of the calibration short by examming the define standard menu
(see "Deflne Standard Menus").
7. Subtract the short offset delay from the value calculated in step 5. This corresponds to the
delay of adapter B.
8. Modify the calibration kit thru definition by entering in the electrical delay of adapter B.
Save this as a user kit.
9. Perform the desired calibration with this new user kit.
10. Measure the test device.