17. To view the conversion loss in the best vertical resolution, press:
Figure 3-9. Conversion LQSS Example Measurement
In this measurement, you set the input power and measured the output power. F'igure 3-9
shows the absolute loss through the mixer versus mixer output frequency. If the mixer
under test contained built-in amplification, then the measurement results would have shown
conversion gain.
M a k i n g
M i x e r
M e a s u r e m e n t s
( O p t i o n
0 8 8

Hide quick links: