HPIB COPY IN PROGRESS, ABORT WITH LOCAL
Error Number An HP-IB copy was already in progress when you requested the HP-IB for
another function. lb abort the first copy, press LLocal), otherwise the HP-IB is
169
unavailable until the ilrst copy is completed.
ILLEGAL UNIT OR VOLUME NUMBER
Error Number The disk unit or volume number set in the analyzer is not valid. Refer to the
46
disk drive operating manual.
Information Continuing with the initialize operation will destroy any data currently on the
Message
disk.
INITIALIZATION FAILED
Error Number The disk initialization failed, probably because the disk is damaged.
47
INSTRUMENT STATE MEMORY CLEARED
Error Number All instrument state registers have been cleared from memory along with any
saved calibration data, memory traces, and calibration kit dehnitions.
56
Additionally, all user-settable selections (such as HP-IB addresses) are set to
their defaults.
INSUFFICIENT MEMORY
Error Number Your last front panel or HP-IB request could not be implemented due to
51
insufficient memory space. In some cases, this is a fatal error from which you
can escape only by presetting the instrument.
Error Number There is not enough memory available for the print or plot function. Increase
the available memory by changing or eliminating a memory-intensive operation
168
such as reducing the number of points in the sweep.
Error ihsafles 1 O-1 3