PROCESSING DISPLAY LIST
Information
The display information is being processed for a screen print to a copy device
and stored in the copy spool buffer. During this time, the analyzer's resources
Message
are dedicated to this task (which takes less than a few seconds.)
PWR MTR NOT ON/CONNECTED OR WRONG ADDRS
Error Number The power meter cannot be accessed by the analyzer. Verify that the power
meter address and model number set in the analyzer match the address and
117
model number of the actual power meter.
REQUESTED DATA NOT CURRENTLY AVAILABLE
Error Number The analyzer does not currently contain the data you have requested. For
30
example, this condition occurs when you request error term arrays and no
calibration is active.
SAVEFAILED. INSUFFICIENTHEMORY
Error Number You cannot store an instrument state in an internal register due to insufficient
151
memory. Increase the available memory by clearing one or more save/recall
registers and pressing m, or by storing files to a disk.
SELECTED SEQUENCE IS EMPTY
Error Number The sequence you attempted to run does not contain instrument commands.
124
Service Error Internal test #'n has failed. Several internal test routines are executed at
Number 112 instrument preset. The analyzer reports the hrst failure detected. Refer to the
HP 8719D/.ZOD/ZZD Network An.algz.er
information on internal tests and the self-diagnose feature.
SEQUENCE ABORTED
Error Number The sequence running was stopped prematurely when you pressed the a
157
key.