The template for component data files is as follows:
! comment line
. . .
where
!
indicates that all following on this Iine is a comment
#
indicates that entries following on this Iine are parameters that are being
specified
frequency
units
parameter
S for S-parameters
format
DB for dB magnitude and angle in degrees
MA for linear magnitude and angle in degrees
RI for real and imaginary pair
Rn
the reference impedance in ohms for the analyzer making the measurement
(R 50 or R 75)
The "format" above is selected by the current selection under the format menu. 'Ib select the
DB format, the format must be LOG MAG. For MA, the format must be LIN MAG, and aII other
format selections wiII output RI data.
Here is an SZP example file for an 21 point measurement of a 20 dB attenuator:
! Network Analyzer HP8720D.06.11 Serial No. US31240052
! <Title line for current channel>
! 23 May 1997 15:26:54
50000000 -56.404 -145.38 -.0083 -.3337 -..0079 -.1606 -58.034 5.0084
7050000000 -54.229
-52.49
16050000000 -50.548 -63.182
-76.33
20050000000 -59.269
4 4 0
Printing, Plotting, and Saving Measurement Results
41.723
119.38
0.0
38.486
70.648
88.746
111.97
103.21
35.461
46.505
78.573
25.793
36.612
19.931
149.78
163.78
39.47
68.46