6. lb measure the standard, when the displayed trace has settled, press:
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.
The analyzer underlines the #p&J softkey after it measures the standard.
7. Disconnect the open, and connect a short circuit to PORT 1.
8. To measure the device, when the displayed trace has settled, press:
The analyzer measures the short circuit and underlines the J%lXW softkey.
9. Disconnect the short, and connect an impedance-matched load to PORT 1. Refer to
"Choosing Calibration Load Standards. n
the softkey corresponding to the load used. If a sliding load is used, press $H,J@@##& to
Otherwise, the message CAUTION: MORE SLIDES NEEDED will be displayed.
11.
12. Repeat the open-short-load measurements described above, but connect the devices in,,,
softkeys. Include any adapters that you would include in your device measurement.
13. To compute the reflection correction coefficients, press:
14. To start the transmission portion of the correction, press:
15. Make a "thru" connection between the points where you will connect your device under
test as shown in Pigure 5-7.
Include any adapters or cables that you will have in the device measurement.
That is, connect the standard device where you will connect your device under
test.