Modifying List Frequencies

List frequencies can be entered or modified using the edit list and edit subsweep menus,
Application of the functions in these menus is described below.

Edit list menu

This menu is used to edit the list of frequency segments (subsweeps) defined with the edit
maximum of 1632 points. The segments do not have to be entered in any particular order:
the analyzer automatically sorts them and shows them on the display in increasing order of
start frequency. This menu determines which entry on the list is to be modifled, while the edit
entry.

Edit subsweep menu

Using the ##EE or $I&; softkey within the edit list menu will display the edit subsweep menu.
This menu lets you select measurement frequencies arbitrarily. Using this menu it is possible to
define the exact frequencies to be measured on a point-by-point basis. For example, the sweep
could include 100 points in a narrow passband, 100 points across a broad stop band, and 50
points across the third harmonic response. The total sweep is deiined with a list of subsweeps
The frequency subsweeps, or segments, can be deihted in any of the following terms:
start/stop/number of points
n
start/stop/step
n
center/span/number of points
n
center/span/step
n
CW frequency
n
The subsweeps can overlap, and do not have to be entered in any particular order. The
analyzer sorts the segments automatically and lists them on the display in order of increasing
start frequency, even if they are entered in center/span format. If duplicate frequencies
exist, the analyzer makes multiple measurements on identical points to maintain the specified
number of points for each subsweep. The data is shown on the display as a single trace that is
a composite of all data taken. The trace may appear uneven because of the distribution of the
data points, but the frequency scale is linear across the total range.
Once the list frequencies have been defined or modified, the list frequency sweep mode can be
The frequency list parameters can also be saved with an instrument state.