Because power meter calibration requires a longer sweep time,~,you may want
Note
to reduce the number of points before pressing ~~kK&~~~~~ ;$IiEI$? . After the
power meter calibration is finished, return the number of points to its original
9. Connect the measurement equipment as shown in Figure 3-l 1.
10. Set the following analyzer parameters:
11. To calibrate the B-channel over the IF range, press:
Once completed, the analyzer should display 0 dBm (-10 dBm, HP 8722D).
12. Make the connections shown in Figure 3-12.
13. Set the Lo source to the desired CW frequency and power level. For this example, the
CW frequency = 1500 MHz
n
source power = 13 dBm
n
3 . 1 4
M a k i n g
M i x e r
M e a s u r e m e n t s
Figure 3-11. Connections for Receiver Chlibration
( O p t i o n
0 8 8