Standard Connections for a Response Error-Correction for Reflection Measurement
8. To measure the standard when the displayed trace has settled, press:
If the calibration kit you selected has a choice between male and female calibration
standards, remember to select the sex that applies to the test port and not the standard.
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.
The analyzer underlines the softkey that you selected after it llnishes the measurement, and
computes the error coefficients.
This calibration allows only one standard to be measured. If you press the
wrong key for a standard, start over with step 6. Lk, not use a thru standard
for a reflection response correction.
You can save or store the measurement correction to use for later
measurements, that use the same measurement parameters. Refer to the
"Printing, Plotting, and Saving Measurement Results" chapter for procedures.
9. This completes the response correction for reflection measurements. You can connect and
measure your device under test.