Frequency Response

Test Limits

ESA Model
Frequency
E4411B &
9 kHz to 1.5 GHz
E4401B
E4402B &
9 kHz to 3 GHz
E4403B
E4404B
9 kHz to 3 GHz
3.01 GHz to 6.7 GHz
E4405B
9 kHz to 3 GHz
3.01 GHz to 6.7 GHz
6.71 GHz to 13.2 GHz
E4407B &
9 kHz to 3 GHz
E4408B
3.01 GHz to 6.7 GHz
6.71 GHz to 13.2 GHz
13.21 GHz to 25 GHz
25 GHz to 26.5 GHz

Test Description

The output of the synthesized sweeper is fed through a power splitter to
a power sensor and the analyzer. The synthesized sweeper's power
level is adjusted at 50 MHz to place the displayed signal at the analyzer
center horizontal graticule line. Measurements are made at various
points depending on the model being tested. The signal source
amplitude is measured with a power meter to eliminate errors due to
source flatness. The power meter is zeroed and calibrated before
starting the measurement.
Chapter 3
Functional Testing
Frequency Response
Minimum
Maximum
(dB)
(dB)
–1.5
–1.5
–1.5
–3.0
–1.5
–3.0
–3.5
–1.5
–3.0
–3.5
–4.0
–4.0
1.5
1.5
1.5
3.0
1.5
3.0
3.5
1.5
3.0
3.5
4.0
4.5