Label Class Menu; Label Kit Menu - Agilent Technologies 8719D User Manual

Network analyzers
Table of Contents

Advertisement

. S22A allows you to enter the standard numbers for the first class required for an Sz2 l-port
calibration. (For default calibration kits, this is the open.)
. S22B allows you to enter the standard numbers for the second class required for an Szz
l-port calibration. (For default calibration kits, this is the short.)
. S22C allows you to enter the standard numbers for the third class required for an Sz2 l-port
calibration. (For default calibration kits, this is the load.)
n
calibration. (For default calibration kits, this is the thru.)
. REV TRANS. allows you to enter the standard numbers for the reverse transmission (thru)
calibration. (For default calibration kits, this is the thru.)
. FWR HAT@4 allows you to enter the standard numbers for the forward match (thru)
calibration. (For default calibration kits, this is the thru.)
. REV MATCH allows you to enter the standard numbers for the reverse match (thru)
calibration. (For default calibration kits, this is the thru.)
n
calibration corrects for frequency response in either reflection or transmission measurements,
depending on the parameter being measured
kits, the standard is either the open or short for reflection measurements, or the thru for
transmission measurements.)
. RESPOIBE B ISflLYH allows you to enter the standard numbers for a response & isolation
calibration. This calibration corrects for frequency response and directivity in reflection
measurements, or frequency response and isolation in transmission measurements.
. TRL TIIlW allows you to enter the standard numbers for a TRL thru calibration.
n
calibration.

Label Class Menu

The label class menus are used to define meaningful labels for the calibration classes. These
then become softkey labels during a measurement calibration. Labels can be up to ten
characters long.

Label Kit Menu

This LA3EL KIT softkey within the modify cal kit menu, provides access to this menu. It
is identi&l to the label class menu and the label standard menu described above. It allows
definition of a label up to eight characters long.
After a new calibration kit has been defined, be sure to specify a label for it. Choose a label
that describes the connector type of the calibration devices. This label will then appear in the
CAL KIT I: ] softkey label in the correction menu and the ?$DX~~ .[
cal kit menu. It will be saved with calibration sets.
a calibration is performed. (For default
when