ANALOG INPUT OVERLOAD
Error Number The power level of the analog input is too high. Reduce the power level of the
60
analog input source.
ANOTHER SYSTEM CONTROLLER ON HP-IB BUS
Error Number You must remove the active controller from the bus or the controller must
37
relinquish the bus before the analyzer can assume the system controller mode.
ASCII: MISSING 'BEGIN' STATEMENT
Error Number The citiiile you just downloaded over the HP-IB or via disk was not properly
193
organized. The analyzer is unable to read the "BEGIN" statement.
ASCII: MISSING 'CITIFILE'STATEMENT
Error Number The citiille you just downloaded over the HP-IB or via disk was not properly
194
organized. The analyzer is unable to read the "CITIFILE" statement.
ASCII: MISSING 'DATA' STATEMENT
Error Number The citillle you just downloaded over the HP-IB or via disk was not properly
195
organized. The analyzer is unable to read the "DATA" statement.
ASCII: MISSING 'VAR' STATEMENT
Error Number The citifile you just downloaded over the HP-IB or via disk was not properly
196
organized. The analyzer is unable to read the "VAR" statement.
AVERAGING INVALID ON NON-RATIO MEASURE
Error Number You cannot use sweep-to-sweep averaging in single-input measurements
13
Sweep-sweep averaging is valid only for ratioed measurements (A/R, B/R, A/B,
and S-parameters). You can use noise reduction techniques, such as narrower
IF bandwidth, for single input measurements.
Error Messages 1 O-3