Saving measurement results into USB memory

You can save measurement results obtained by the E4981A into USB memory
as .CSV files. You can later load your saved files into an application program
running on a PC.
Data buffer 3 is used to perform save data operation through front panel
and web server. Also, it is not possible to load data through web server.
Before saving measurement results into USB memory, you need to complete
some tasks preparatory to initiating measurement so that the data buffer
memory contains the necessary data. For more information, see
measurement results into USB memory:" on page
You cannot load measurement results from USB memory into the E4981A.
Measurement result format
Measurement results are output in the format shown in
The Data A and Data B fields of a measurement result file differ depending on
whether the "effective digits" setting is on or off (controlled by the
:FORMat:ASCii:LONG command).
Table 4-10
Measurement result format
Measurement
Screen
one point
one point
Data A
Outputs the measurement data for the primary parameter.
The Data A field uses the following two different fixed-length ASCII formats:
When the "effective digits"
setting is off:
(:FORMat:ASCii:LONG OFF)
When the "effective digits"
setting is on:
(:FORMat:ASCii:LONG ON)
86
Comparator
Resul t Format
OFF
<Data A>,<Data B>,<Status>
ON
<Data A>,<Data B>,<Status>,<BIN No.>
SN.NNNNNESNN
(S:+/-, N:0 to 5, E: exponent character)
SN.NNNNNNNNNESNN
(S:+/-, N:0 to 9, E: exponent character)
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter