Viewing Analysis Results In The Graph Tab; Editing Formulas And Constants - Keithley 4200-SCS Reference Manual

Semiconductor characterization system
Hide thumbs Also See for 4200-SCS:
Table of Contents

Advertisement

Model 4200-SCS Reference Manual

Viewing analysis results in the Graph tab

If a new column (vector) is added to the Sheet tab Data worksheet after creating/modifying a
formula, it can be plotted in the Graph tab, just as any other column (vector). See
Figure 6-368
Added linear regression line to the graph shown in
NOTE

Editing formulas and constants

To Edit a Formulator formula, do as follows:
1.
4200-901-01 Rev. S / May 2017
before the first row, a valid calculation is not possible for the first row.
Therefore, the Formulator returns the #REF notation in the first row.
A column will contain multiple instances of #REF if the Formulator function requires
multiple prior cells for the calculation. For example, if the MAVG function is using five data
points to calculate a moving average of a column containing five values, the first two and
last two cells will contain #REF.
Added linear regression line, calculated through the formula
PLATEAULINE = REGFIT(COLLECTORV,COLLECTORI,52,152)
For information about using the Graph tab, refer to
using the Sheet tab.
In the lower Formula box double-click the formula that you wish to edit. A copy of the
formula appears in the upper Formula box. See the example in
Return to
Section Topics
Section 6: Keithley Interactive Test Environment (KITE)
Figure
6-361.
Displaying and analyzing data
Figure
Figure
6-368.
6-369.
6-315

Advertisement

Table of Contents
loading

Table of Contents