Keithley 4200-SCS Reference Manual page 900

Semiconductor characterization system
Hide thumbs Also See for 4200-SCS:
Table of Contents

Advertisement

Section 11: Pulse Source-Measure Concepts
Scope card ext trig
For pulse-measure synchronization, the scope must be set for the normal sweep mode, and set to
be triggered by the leading-edge or falling edge trigger from the 4205-PG2 pulse generator. Refer
to the ZTEC scope User's Manual for details on triggering. Triggers from the pulse generator are to
be applied to the EXT TRIG connector of the scope card.
Trigger connections: pulse-measure
As shown
scope card. The SMA cable and adapter for the scope card are supplied items.
Figure 11-16
Trigger connections for pulse-measure synchronization
Example LPT function sequence: pulse-measure synchronization
The following LPT function sequence uses the software trigger to initiate a 3-pulse burst for
CHANNEL 1. The three output pulses will trigger the scope to perform three measurements. It
assumes the scope is configured to trigger on leading edge triggers from the pulse generator:
// Stop pulse generator output:
pulse_halt(VPU1);
// Turn off trigger output:
pulse_trig_output(VPU1, 0);
// Set the output trigger polarity to positive:
pulse_trig_polarity(VPU1, 1);
// Select Software trigger source:
pulse_trig_source(VPU1, 0);
// Set channel 1 for a burst count of 3:
pulse_burst_count(VPU1, 1, 3);
// Turn channel 1 on:
pulse_output(VPU1, 1, 1);
11-32
Figure
11-16, TRIGGER OUT of the pulse generator card is connected to Trg Ext of the
Model
4200-SCP2HR
4205-PG2
or 4200-SCP2
Pulse Generator
KEITHLEY
4205
PG2
TRIGGER
OUT
TRIGGER
OUT
TRIGGER
IN
CHANNEL 1
CHANNEL 2
White SMA Cable
(15cm, 6in)
(male-to-male)
Return to
Model
Scope
KEITHLEY
4200
SCP2HR
CHANNEL 1
EXT
TRG
EXT
TRG
EXT
CLK
= SMB Male to SMA Female Adapter