Model 4200-SCS Reference Manual
Remarks
Table 8-22
Available current ranges
4200-901-01 Rev. S / May 2017
Vrange_type
Vrange
Irange_type
Irange
For the PMU, use this function to set the voltage source range for pulse output, set the
voltage and current measure range types, and set the actual voltage and current
measure ranges. For the PGU, use this function to set the voltage source range for
pulse output.
Measure range types for the PMU:
• Fixed: Use this range type to specify a fixed measure range (Vrange or Irange).
• Limited Auto: Select this range type to use the fixed measure as the lowest range
that will be used for automatic ranging.
• Auto: Use this range type to automatically select the optimum measure range. The
specified fixed measure range (Vrange or Irange) is not used in autorange
mode, but must be a valid range.
• The current ranges available depend on the source range and whether the system
includes a 4225-RPM, as shown in the next table.
Current measure range (A)
0.8
0.2
0.01
0.001
0.0001
0.00001
0.000001
0.0000001
Auto or limited autoranging is available only when using the advanced mode in the
pulse_exec
function. Ranging is controlled per-channel and may be combined with load
Return to
Section Topics
Section 8: Keithley User Library Tool (KULT)
Voltage measure range type (PMU):
0 = Auto
1 = Limited auto
2 = Fixed
Vrange_type parameter is ignored by the PGU. Autorange
(0) and limited autorange (1) are not valid for the Segment
ARB pulse mode.
Voltage measure range (PMU)
Vrange parameter is ignored by the PGU and is ignored by
the PMU when autorange is selected.
Current measure range type (PMU):
0 = Auto
1 = Limited auto
2 = Fixed
Irange_type parameter is ignored by the PGU. Autorange
(0) and limited autorange (1) are not valid for the Segment
ARB pulse mode.
Current measure range (PMU)
Irange parameter is ignored by the PGU and is ignored by
the PMU when autorange is selected.
PMU source range (V)
n/a
10
10
n/a
n/a
n/a
n/a
n/a
RPM source range (V)
40
n/a
10 or 40
10
10 or 40
10
10
10