Pulse Generator Connections; Scope Connections - Keithley 4200-SCS Reference Manual

Semiconductor characterization system
Hide thumbs Also See for 4200-SCS:
Table of Contents

Advertisement

Section 11: Pulse Source-Measure Concepts

Pulse generator connections

Figure 11-22
Figure 11-22
Basic pulse generator connections
Pulse Generator Card
SLOT
SLOT
8
7
KEITHLEY
KEITHLEY
4205
4200
SCP2HR
PG2
TRIGGER
CHANNEL 1
OUT
EXT
TRG
TRIGGER
IN
EXT
CLK
CHANNEL 1
CHANNEL 2
CHANNEL 2
NOTE Use supplied torque wrench to tighten

Scope connections

Figure 11-23
NOTE
Figure 11-23
Basic scope connections
SLOT
SLOT
8
7
KEITHLEY
KEITHLEY
4205
4200
PG2
SCP2HR
CHANNEL 1
CHANNEL 1
EXT
TRG
TRIGGER
OUT
EXT
CLK
CHANNEL 2
CHANNEL 2
TRIGGER
IN
11-40
shows a system that uses basic 2-channel pulse generator connections to DUTs.
Model 4200-SCS
INSTRUMENTS
SLOT
SLOT
SLOT
SLOT
SLOT
SLOT
6
5
4
3
2
KEITHLEY
KEITHLEY
KEITHLEY
KEITHLEY
4210
4210
4200
4200
SMU
SMU
SMU
SMU
SENSE LO
SENSE LO
SENSE LO
SENSE LO
SENSE
SENSE
SENSE
SENSE
FORCE
FORCE
FORCE
FORCE
PA CNTRL
PA CNTRL
PA CNTRL
PA CNTRL
SMA connections to
8 inch-lbs.
shows basic 2-channel scope card connections to the signals to be measured.
The BNC cable supplied with the scope card is a 50 Ω cable. When using this cable
with the scope be sure that the scope input is set to the 50 Ω mode. Failure to do so
could cause incorrect measurements.
Scope Card
Model 4200-SCS
INSTRUMENTS
SLOT
SLOT
SLOT
SLOT
SLOT
SLOT
6
5
4
3
2
KEITHLEY
KEITHLEY
KEITHLEY
KEITHLEY
4210
4210
4200
4200
SMU
SMU
SMU
SMU
SENSE LO
SENSE LO
SENSE LO
SENSE LO
SENSE
SENSE
SENSE
SENSE
FORCE
FORCE
FORCE
FORCE
PA CNTRL
PA CNTRL
PA CNTRL
PA CNTRL
Return to
Equivalent Circuit (each channel):
4200-SCS
Pulse
Generator
Card
1
SMA
Pulse Output
(Channel 1)
Common
SMA
Pulse Output
(Channel 2)
Common
White SMA Cable (2m, 6ft)
(male-to-male)
1 of 2
Equivalent Circuit (each channel):
4200-SCS
Scope
DSO
Card
1
BNC
Scope Input
Channel 1
Common
Black BNC Cables
(2m, 6ft)
(male-to-male)
BNC
Scope Input
Channel 2
Common
Section Topics
Model 4200-SCS Reference Manual
DUT
DUT
DUT
Waveform to
be measured
Waveform to
be measured
Waveform to
be measured
4200-901-01 Rev. S / May 2017

Advertisement