Keithley 4200-SCS Reference Manual page 709

Semiconductor characterization system
Hide thumbs Also See for 4200-SCS:
Table of Contents

Advertisement

Model 4200-SCS Reference Manual
Remarks
4200-901-01 Rev. S / May 2017
pulse_meas_timing
int pulse_meas_timing(INSTR_ID instr_id, int Chan, double,
StartPercent, double StopPercent, int NumPulses);
instr_id
chan
StartPercent
StopPercent
NumPulses
Use the
pulse_sample_rate
NOTE
measurements.
There are four types of pulse measurements that are performed by the Model
4225-PMU: Spot mean discrete, spot mean average, waveform discrete, and waveform
average. Use the following pulse generator functions to configure pulse measurements:
• Use the pulse_meas_sm function to configure spot mean measurements; select the
data acquisition type, set the readings to be returned, enable or disable time stamp,
and set LLEC. See
Spot mean measurements
• Use the pulse_meas_wfm function to configure waveform measurements; select the
data acquisition type, set the readings to be returned, enable or disable time stamp,
and set LLEC. See
Waveform measurements
• Use the pulse_meas_timing function to set measurement timing. For spot mean
measurements, portions of the amplitude and base levels are specified for sampling.
For pre-data and post-data waveform measurements, a percentage of the entire
pulse duration is specified. See
measurement timing.
Before calling the pulse_meas_timing function, use the pulse_meas_sm
NOTE
or pulse_meas_wfm function to configure the measurement type.
Measurement types
There are two types of pulse measurements: Spot mean and waveform. The
pulse_meas_sm
function is used to configure
pulse_meas_wfm
function is used to configure
Return to
Section Topics
Section 8: Keithley User Library Tool (KULT)
provides its own analog LLEC (assuming a short cable from
the RPM to the DUT).
Instrument ID: PMU1, PMU2, and so on.
PMU channel: 1 or 2
Spot mean: Start location for spot mean measurements,
specified as a percentage of the widths for
the amplitude and base level (see
Waveform: Pre-data for the amplitude, specified as a
percentage of the amplitude pulse duration (see
Figure
8-120).
Spot mean: Stop location of the spot mean measurements,
specified as a percentage of the widths for
the amplitude and base level (see
Waveform: Post-data for the amplitude, specified as a
percentage of the amplitude pulse duration (see
Figure
8-120).
Number of pulses to output and measure (1 to 10,000).
function to set the sampling rate for pulse
for details.
for details.
Measurement timing
Spot mean measurements
Waveform
Figure
Figure
for details on pulse
and the
measurements.
8-116).
8-116).
8-193