XN command: Configure graph X axis for electrical parameter
The following command string configures the X axis of the graph to plot an electrical parameter:
XN 'AAAAAA', B, ±CCCC.CCC, ±DDDD.DDD
XN
X axis SMU channel name
AAAAAA
X axis scale type
B
X axis minimum value
±CCCC.CCC
X axis maximum value
±DDDD.DDD
Example
The command string:
XN 'V1', 1, -5, 5
does the following:
XT command: Configure graph X axis for time domain
The following command string configures the X axis of the graph to plot time domain values (in
seconds):
XT AAAA.AA, BBBB.BB
XT
X-axis minimum time value
AAAA.AA
X-axis maximum time value
BBBB.BB
Example
The command string:
XT 0, 10
does the following:
4200-901-01 Rev. S / May 2017
=
Command string prefix
=
The SMU channel name for the X axis, up to 6 characters long. Must be
one of the SMU channel names that you specify on the channel definition
(DE) page.
= 1 Linear scale
= 2 Log scale
=
±9999 (volts)
=
±999 (amps)
=
±9999 (volts)
=
±999 (amps)
Specifies that values from SMU channel V1 are to be plotted on the X axis.
Sets up the X axis to be scaled linearly between -5 V and +5 V.
=
Command string prefix
=
0.01 to 9999 (seconds)
=
0.01 to 9999 (seconds)
Specifies that time domain values are to be plotted on the X axis.
Sets up the X axis to be scaled between 0 and 10 s.