Figure 4-12
Interlock connector location
Interlock cables
Use the supplied interlock cable (236-ILC-3) or the equivalent to make interlock connections.
Typical interlock connections
Figure 4-13
component test fixture. The test fixture has a safety interlock switch connected to its lid. When the
lid is closed, the interlock circuit is closed (asserted), and SMU ±200 V ranges are enabled.
Conversely, the interlock circuit is open (deasserted) when the lid is open, and SMU ±200 V
ranges are disabled. A safety interlock cable is supplied with the 4200-SCS, allowing it to directly
interface to the interlock circuits of the Keithley Instruments Models 8006 and 8007 test fixtures.
Figure 4-13
Typical interlock connections
4200-901-01 Rev. S / May 2017
Interlock
Connector
Trigger Master
shows typical interlock connections. In this example, the 4200-SCS is connected to a
Interlock Connector
Interlock Connector
INTERLOCK
WIRING
1
!
LID
INTERLOCK
8
1
11
KEITHLEY
V
Return to
1
F
1
F
8
7
6
5
1
F
F
4200-TM
Trigger Link
Connectors
(Not Currently
Supported)
Model 4200-SCS
1
F
1
F
8
7
Interlock Cable
Model 8006
TRIAX
!
BNC
V
6
V
11
V
1
7
3 V
BNC 2
BNC 1
!
WARNING
V
1
F
V
18
V
V
F
Section Topics
Section 4: Connections and Configuration
3
1
1
F
F
6
5
3
1
1
1
F
F
F
F
BINDING POST
3
1
3
1
6
5
4-21

Advertisement

Table of Contents
loading

Table of Contents