Section 9: Keithley External Control Interface (KXCI)
Table 9-2 (continued)
System mode commands
1
Page
Command Function
SM
YA
Configure graph Y1
(cont.)
axis
YB
Configure graph Y2
axis
MD
ME
Control measurements
6
Any
DO
Obtain output data
SR
Fixed source ranging
9-16
Command String
YA 'AAAAAA', B, ±CCCC.CCC, ±DDDD.DDD
AAAAAA= Axis name (an SMU channel that is specified on DE page)
B = 1 Linear scale
= 2 Log scale
= 3 Log scale absolute value
±CCCC.CCC = ±9999 volts Minimum value
= ±999 amps
±DDDD.DDD = ±9999 volts Maximum value
=±999 amps
YB 'AAAAAA', B, ±CCCC.CCC, ±DDDD.DDD
AAAAAA= Axis name (SMU channel that is specified on DE page)
B = 1 Linear scale
= 2 Log scale
= 3 Log scale absolute value
±CCCC.CCC = ±9999 volts Minimum value
= ±999 amps
±DDDD.DDD = ±9999 volts Maximum value
= ±999 amps
MEA
A = 1 Trigger test, store readings in cleared buffer
= 2 Trigger test, store readings in cleared buffer
= 3 Trigger test, append readings to buffer
= 4 Abort test
DO 'AAAAAA'
AAAAAA = Name of measurement channel
DO 'CHnT'where: n = Absolute channel number
SR A,B
A = The channel controlled: a value between 1 and the number of channels
in the system (8 max).
B = 0
Auto
= 1
Best fixed
> 0 to 1.0 Fixed range