Model 4200-SCS Reference Manual
Table 9-12
KXCI CVU command strings
KXCI CVU commands
User Mode, System Mode, Modeless
User Mode
KXCI command: :CVU:MEASZ?
Purpose: Triggers and returns single Z-measurement using current CVU settings. When the command
System Mode
KXCI command: :CVU:SOAK:DCV voltage
Purpose: Sets the presoak DC voltage for CVU card for ALL sweeps (that is, frequency sweep, DC
Parameters: voltage: Voltage to bias before test sequence begins. Voltages range from -30 V to +30 V.
KXCI command: :CVU:BIAS:DCV:SAMPLE biasv, samples
Purpose: Configures the CVU to bias a DC voltage and sample n Z-measurements for the CVU card.
Parameters: biasv: Voltage to source while sampling Z-measurements. Voltages range from -30 V to +30
KXCI command: :CVU:SWEEP:FREQ fstart, fstop, <order>
Purpose: Configures the CVU to sweep frequency and sample Z-measurements for the selected CVU
Parameters: fstart: The frequency that is used for capturing the initial sample in the sweep.
KXCI command: :CVU:SWEEP:DCV dcvstart, dcvstop, dcvstep
Purpose: Configures the CVU to sweep DC voltage and sample Z-measurements for the selected CVU
Parameters: dcvstart: start voltage
4200-901-01 Rev. S / May 2017
is complete, the single reading is available over GPIB or Ethernet.
voltage sweep, AC voltage sweep, and so on) for the selected CVU card.
The other parameters are set by their respective commands.
n represents the integer value passed to the samples parameter.
V.
samples: The number of Z-measurements the CVU will make for the test operation. Valid
values range from 1 to 4096.
card. Only the start frequency, stop frequency, and sweep order are set here. The other
parameters are set by their respective commands.
fstop: The frequency that is used for capturing the final sample in the sweep.
order: <optional parameter> Determines whether the test will bias DC voltage and sweep
frequency, or step voltage and sweep frequency.
1: The voltage will just bias the voltage set by the :CVU:DCV command.
the behavior when the parameter is omitted.
2: The voltage will step based on the vstart, vstop, vstep parameters set
:CVU:SWEEP:DCV command.
NOTE
card.
dcvstop: stop voltage
dcvstep: voltage step size
NOTE
Return to
Section 9: Keithley External Control Interface (KXCI)
Values are coerced to one of the 28 discrete frequencies or 37
discrete frequencies for the 4200-CVU and 4210-CVU, respectively.
The other parameters (AC voltage level, frequency, and so on) are
set by their respective commands. Valid voltage levels are -30 V to
+30 V.
Section Topics
This will also be
by the
9-97