Section 6: Keithley Interactive Test Environment (KITE)
Compliance: The Compliance parameter edit box specifies any valid SMU voltage
compliance limit, the units for which are selected through the combo box at the right of the
edit box.
Power On Delay: When an ITM test is run, the SMUs for the given test power-on in a
specific sequence. A power-on delay can be set between the SMUs in the test. The Power
On Delay field is used to set this delay, which can be set from 0 to 0.100 s. For details, see
Power On
Pulse Mode: When selected, the Pulse Mode combo box allows definition of a current
pulse list sweep. The on and off times can be set from 5 ms to 10 s. Pulse output goes to
the sweep list levels during the pulse on times. During the off times, pulse output returns to
the specified Base Current level. For details, see
NOTE
Understanding and configuring the Voltage List Sweep forcing function
The Voltage List Sweep forcing function increments through a series of voltage values, each
of which is individually user-specified. Almost any pattern of steps may be configured, subject to
the range limitations of the SMU and the specified current compliance. Refer to the illustrations in
Understanding List Sweeps in
Options window is shown in
Figure 6-139
Voltage List Sweep Forcing Functions/Measure Options window
6-116
Delay.
Pulse Mode can be selected ONLY when source and measure ranges are fixed. In
other words, Pulse Mode is disabled if the source or measure range is set to AUTO.
general. A typical Voltage Sweep Forcing Functions/Measure
Figure
Return to
Pulse
6-139.
Parameter list
Section Topics
Model 4200-SCS Reference Manual
Mode.
4200-901-01 Rev. S / May 2017