Output Data Formats; Data Format For System Mode Readings; Data Format For User Mode Readings - Keithley 4200-SCS Reference Manual

Semiconductor characterization system
Hide thumbs Also See for 4200-SCS:
Table of Contents

Advertisement

Section 9: Keithley External Control Interface (KXCI)
Table 9-8
Source Setup (Page SS) KXCI defaults
Name
Sweep mode
Start
Stop
Step
Number of steps
Compliance
Constant
V1
V4
VS1
VS2

Output data formats

Data format for system mode readings

For system mode operation, the DO command is used to obtain one or more triggered readings.
After sending the DO command string and addressing the Model 4200-SCS to talk, the output data
string will be sent to the computer in the following format:
X±N.NNNNE±NN,X±N.NNNNE±NN,...X±N.NNNNE±NN
Data status
X =
Reading (mantissa and exponent)
±N.NNNN E±NN

Data format for user mode readings

For user mode operation, the TI or TV command string is used to trigger and obtain a reading.
After sending the TI or TV command string and addressing the Model 4200-SCS to talk, the output
data string will be sent to the computer in the following format:
32. The hierarchy for data status is L, V, X, C, T, N.
33. Scientific notation for the reading exponent:
E+00=10
E-03=10
E-06=10
E-09=10
E-12=10
9-50
VAR1
(Command VR)
V3
Linear
0.00 V
1.00 V
0.010 V
101
100.0 mA
Source
COM
0.00 V
V
0.00 V
V
0.00 V
V
0.00 V
32
N
Normal
L
Interval too short
V
Overflow reading (A/D converter saturated)
X
Oscillation
C
This channel in compliance
T
Other channel in compliance
0
-3
(m)
-6
(µ)
-9
(n)
-12
(p)
Return to
VAR2
(Command IP)
I2
Linear
20.00 µA
N/A
20.00 µA
5
2V
Compliance
Command
105.0 mA
N/A
100.0 mA
VC
N/A
SC
N/A
SC
33