Keithley 4200-SCS Reference Manual page 220

Semiconductor characterization system
Hide thumbs Also See for 4200-SCS:
Table of Contents

Advertisement

Section 6: Keithley Interactive Test Environment (KITE)
Table 6-6
Forcing function summary
Category Name
Static
Open
Common
Current Bias Maintains a selected constant-current state at the terminal, subject to the user-
Voltage Bias Maintains a selected constant-voltage state at the terminal, subject to a user-
Sweep
Current
Sweep
Voltage
Sweep
List sweep Current
List Sweep
Voltage
List Sweep
Step
Current Step Increments a current to two or more levels, each of which is held constant during
Voltage Step Increments a voltage to two or more levels, each of which is held constant during
For detailed information about each forcing function, refer to
parameters area later in this
6-94
Description
Maintains a zero-current state at the terminal, subject to the maximum voltage
compliance of the connected SMU.
Maintains a zero-voltage state at the terminal, subject to the maximum current
compliance of the connected SMU.
specified voltage compliance for the connected SMU.
specified current compliance of the connected SMU.
Increments a series of current values at a rate that is determined by the Timing
and Speed settings in the ITM Definition tab. Generates parametric curve data
that is recorded in the Sheet tab Data worksheet for the ITM and can be plotted in
the ITM Graph tab. A dual current sweep can also be performed by an SMU (see
Understanding a Dual Sweep later in this
Increments a series of voltage values at a rate that is determined by the Timing
and Speed settings in the ITM Definition tab. Generates parametric curve data
that is recorded in the ITM Sheet tab Data worksheet and can be plotted in the ITM
Graph tab. A dual voltage sweep can also be performed by an SMU (see
Understanding a Dual Sweep later in this
Steps through a list of user-specified current values, at a rate that is determined by
the Timing and Speed settings in the ITM Definition tab. Generates parametric
data that is recorded in the ITM Sheet tab Data worksheet and can be plotted in
the ITM Graph tab, if appropriate.
Steps through a list of user-specified voltage values, at a rate that is determined by
the Timing and Speed settings in the ITM Definition tab. Generates parametric
data that is recorded in the ITM Sheet tab Data worksheet and can be plotted in
the ITM Graph tab, if appropriate.
the progress of a Current Sweep, a Voltage Sweep, a Current List Sweep, or a
Voltage List Sweep at another terminal. For each Current Step level, parametric
curve data is recorded in the ITM Sheet tab Data worksheet. The combined data
can be plotted in the ITM Graph tab, resulting in a series (family) of curves.
the progress of a Current Sweep, a Voltage Sweep, a Current List Sweep, or a
Voltage List Sweep at another terminal. For each Voltage Step level, parametric
curve data is recorded in the ITM Sheet tab Data worksheet. The combined data
can be plotted in the ITM Graph tab, resulting in a series (family) of curves.
section.