Keithley 4200-SCS Reference Manual page 946

Semiconductor characterization system
Hide thumbs Also See for 4200-SCS:
Table of Contents

Advertisement

Section 11: Pulse Source-Measure Concepts
DrainCompliance
NPLC
SweepDelay
MaxIg
MaxId
MaxPowerGate
MaxPowerDrain
GateSMU
DrainSMU
GateVPU
DrainVPUHigh
DrainVPULow
DrainVSize
DrainISize
GateVSize
GateISize
PostData
11-78
the value given is the minimum measurement range used, with automatic
ranging for larger currents.
1
Full Auto
2
Limited Auto 10 pA
3
Limited Auto 100 pA
4
Limited Auto 1 nA
5
Limited Auto 10 nA
6
Limited Auto 100 nA
7
Limited Auto 1 uA
8
Limited Auto 10 uA
9
Limited Auto 100 uA
10 Limited Auto 1 mA
11 Limited Auto 10 mA
12 Limited Auto 100 mA
(double) The current compliance for the DUT
Drain.
(double) The DC measurement integration time
in NPLC (Number of Power Line cycles).
(double) Time, in seconds, between the DC source and measure for each
sweep point.
(double) The max current allowed on the DUT Gate. If the Gate current
becomes greater than MaxIg, the test will exit.
(double) The max current allowed on the DUT Drain. If the Drain current
becomes greater than MaxId, the test will exit.
(double) The max power allowed to the DUT Gate. If the gate power becomes
greater than the MaxPowerGate, the test will exit.
(double) The max power allowed on the DUT Drain. If the Drain power
becomes greater than the MaxPowerDrain, the test will exit.
(char*) String representing the ID for the SMU connected to the DUT Gate
terminal.
(char*) String representing the ID for the SMU connected to the DUT Drain
terminal.
(char*) String representing the ID for the PG2 connected to the DUT Gate
terminal.
(char*) String representing the top VPU on the Model 4205-PCU. This is the
card on the right (or the card with the lower ID number), when facing the back
of the system.
(char*) String representing the bottom VPU on the Model 4205-PCU. This is
the card on the left (or the card with the higher ID number), when facing the
back of the system.
(int) Sizes of the output arrays. All arrays should be the same size
and need to be large enough to hold all sweep points.
(int) Turns on real time graphing (1 = Real Time Graphing, 0 = No Real Time
Graphing).
Return to
Section Topics
Model 4200-SCS Reference Manual
4200-901-01 Rev. S / May 2017

Advertisement

Table of Contents
loading

Table of Contents