Model 4200-SCS Reference Manual
Figure 9-11
Sweep resulting from the VR1, 1, 5, 1, 0.01 command string
VP and IP commands: VAR2 setup
If the source that is being used is an SMU, the source mode for the VAR2 steps can be voltage or
current. If, however, a voltage source (VS1...VS8) is being used, the source mode must be voltage
(for configuration details, refer to
Parameters for the VAR2 step function include the start value, step value, and the number of
steps. In
Figure 9-12
Steps resulting from the VP 5, 5, 3, 0.01 command string
4200-901-01 Rev. S / May 2017
´
= Measurement
Figure 9-12
the VAR2 step function starts at 5 V, steps in 5 V increments and has 3 steps.
15V
10V
5V
0V
´
= Measurement
Return to
Section Topics
Section 9: Keithley External Control Interface (KXCI)
´
5V
´
4V
´
3V
´
2V
´
1V
0V
Section
7).
9-27