Stopping a Loading, PCA or Bolus Dose

CHANNEL SELECT
1.
Press
STOP LOAD, STOP PCA
2.
Press
applicable.
To stop dose and resume current program, press Yes soft
3.
key.
NOTE:
Available soft key and stop confirmation screen are dependent
on the type of dose currently infusing and current infusion mode.

Changing Programming Parameters During an Infusion

CHANNEL SELECT
1.
Press
PROGRAM
2.
Press
3.
Set key to program position or if Authorization Code is
enabled, enter
4
CHANGE MODE
4.
Press
5.
Select desired infusion mode.
6.
Continue programming. See applicable procedure:
Setting Up PCA Dose Only
Setting Up Continuous Infusion Only
Setting Up PCA + Continuous Infusion
3-22
Programming
Infusion Modes (Continued)
key.
STOP BOLUS
or
key.
soft key.
-digit code.
soft key.
soft key as
®
Alaris
Summary
A
Morphine
Loading
1
mg
Dose:
[Conc]: 1 mg/mL
>Press START to Close
Summary
STOP
START
LOAD
Guardrails Drug Setup
A
IMS Pre-fill 30mL
2
PCA
mg
DOSE
LOCKOUT
20
minutes
INTERVAL
CONT
2
mg/h
DOSE
MAX
25
mg/4 h
LIMIT
[Conc]: 1 mg/mL
>Press CONFIRM
CHANGE
CONFIRM
MODE
System (with v9 Model 8000) DFU
PCA Module Section