Configuration
Automatic Desflurane Compensation
If Desflurane concentration data is communicated
to the Fabius plus by an external agent analyzer,
the following occurs:
1 Des auto appears in the status bar at the top of
the screen
2 the »Des Comp On/Off« soft key label is remo-
ved
the Fabius plus will automatically perform the
corresponding flow sensor compensation.
Automatic Desflurane compensation always overri-
des the functionality of the Desflurane compensa-
tion soft key.
If communication is disconnected or lost between
the Fabius plus and an external agent analyzer
while Desflurane is in use:
the Des auto message is removed from the sta-
tus bar
the »Des Comp On/Off« soft key label reap-
pears as »Des Comp Off«.
To maintain Desflurane compensation, manually
activate Desflurane compensation to ensure accu-
rate volume measurements.
148
1
2
Instructions for Use Fabius plus SW 3.n