Switching the volume alarms on and off

Volume Control
A
1 Press the VolumeAlarms ON/OFF softkey (A).
The key label changes from Volume Alarms
ON to Volume Alarms OFF.
Instead of the upper and lower alarm limits, the
symbol appears, indicating the alarm is
deactivated.
The volume alarms are deactivated.
NOTE
The function Volume Alarms ON /OFF is
available in the standard view of the ManSpont
mode. If the
key is pressed in ManSpont
mode, the softkey Volume Alarms ON /OFF is
not displayed.
Instructions for use Fabius GS premium SW 3.n
Automatic setting of the pressure
threshold
Volume Control
A
1 Press the AutoSet softkey (A).
The pressure threshold for the peak pressure
(PEAK) is set to 4 cmH
current plateau pressure (PLAT).
NOTE
The pressure threshold must not be below
5 cmH
O (5 hPa) or above 30 cmH
2
NOTE
If no current measured value is available for the
plateau pressure (PLAT), pressing the softkey
has no effect.
NOTE
In SIMV/PS mode, the pressure threshold
depends on the pressure of the mandatory
breaths.
Configuration
O (hPa) below the
2
O (30 hPa).
2
145