Alarms
Downgrading and acknowledging alarm
messages
Some alarms can be downgraded to low priority
(note) or they can be cleared completely. The
relevant alarms can be recognized in the table
''Alarm – Cause – Remedy'' on page 162 by the
following remedial messages:
Remedial message
Use "ALARM RESET"
to downgrade alarm
priority.
Use "ALARM RESET"
to acknowledge alarm.
There are 2 options for downgrading or resetting
the alarms:
Option 1:
A
Or
Option 2:
Alarms
B
118
Effect
Alarm priority is
changed to low (note).
Alarm is reset.
C
Option 1
Touch the ALARM
RESET button (A) in the
header bar and confirm.
All the alarms displayed
in the alarm message
field will be downgraded
or reset.
After an alarm is downgraded, the respective
priority rating is maintained, see table ''Alarm –
Cause – Remedy'' on page 162.

Opening the alarm history

The alarm history records all alarm messages for
the current case in chronological sequence.
1 Open the Alarms dialog window.
2 Touch the Alarm history tab (A).
Alarms
A
Use the rotary knob or the arrow buttons to scroll
the cursor up (B) or down (C).
The alarm history is deleted when Perseus is shut
down or a new case is started.
Instructions for use Perseus A500 SW 1.1n
Option 2
In the Alarms > Cur-
rent alarms dialog win-
dow (B), touch the
Reset all button (C) and
confirm.
All alarms will be down-
graded or reset.
B
C