Automatic Self Test; Test History - Dräger Zeus Infinity Empowered Instructions For Use Manual

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Startup

Automatic self test

Touch the Auto self test tab (A).
Test details
A
Time-controlled self test
starts at
Time-controlled self test
The automatic self test is set as standard in the
general basic settings, see page 336.
The default values can be changed for the next test:
– Time-controlled self test starts at (B)
– Time-controlled self test (C) On/Off
The new values apply to the next automatic self
test. After the next automatic self test is completed,
the values are reset to the default values.
Factory setting: Off
When activating On, the following messages ap-
pear:
– Auto self test is enabled. Conduct Pretest.
Check start time.
When activating Off, the following message ap-
pears:
– Auto self test is disabled.
102
B
C

Test history

Calling up the test history:
Touch the Test history tab (A).
Test details
The results of previous tests are displayed with the
following information:
– Date and Time
– Test
– Name of tester
– Number of test loops
– Result
If a test did not detect any faults but is incomplete,
the result is indicated with canceled.
Instructions for use Zeus Infinity Empowered SW 2.n
A