4 General operating procedures
4.2 Updating the firmware
70

Option 2: Firmware update via the network

4.2.2.2
Option 2: Firmware update via the network
Opening the web dialog
1. In SIDEXIS Manager, start the
2. Click on the
A password dialog box opens.
3. In the field
4. In the field
"Facescan Device Service "
The
Select Update dialog
Select Update dialog
➢ In the menu bar, select the menu item
"Facescan Firmware Update"
The
Starting the update
"Enter Update Menu"
1. Press the
A further
"Browse..."
2. Press the
3. Navigate to the firmware update file and select it.
4. Press the
"Upload Image"
"Run Update"
5. Press the
The update starts.
The update ends automatically after around 5 minutes with a text
message.
"Reboot"
6. Press the
7. Wait around a minute until the green LED on the Facescan unit
begins to light up.
8. Only if the SIRONA browser is not being used:
Refresh the browser display (e.g. in Windows® Internet Explorer:
press [F5]).
The update is completed.
Concluding the update
1. Switch the unit on.
2. Only at initial installation: Perform a complete unit calibration [ → 155].
3. Only at initial installation: Perform a white balance [ → 190].
"Facescan Configuration"
"Facescanner in browser"
"User"
"service"
enter
.
"Password"
"sirona"
enter
web dialog opens.
window opens.
button.
"Facescan Firmware Update"
button.
button.
button.
button.
Sirona Dental Systems GmbH
Service Manual GALILEOS
application.
button.
.
"UPDATE"
.
window opens.
61 25 665 D3437
D3437.076.01.15.02 06.2012