Setting Up The X-Ray Component Via The Sixabcon Program - Sirona Orthophos XG 3D/Ceph Installation Manual

Table of Contents

Advertisement

11 Initial startup
11.7 Setting up the X-ray component via the SIXABCon program
104
11.7
Setting up the X-ray component via the
SIXABCon program
Since it is addressable via the network, the X-ray component can in
principle be activated for X-ray image acquisition by any of the PCs
connected to the network.
The program SiXABCon is used to manage networkable X-ray
components. It can be executed on any PC in the network on which
SIDEXIS has been installed.
IMPORTANT! In order to avoid IP address conflicts, you should never
operate several networkable X-ray components in the network using the
same IP address. Each X-ray component needs to be assigned a unique
IP address.
Factory setting of the TCP/IP address of the unit:
192.168.15.240 (subnet mask: 255.255.255.0)
Checking existing IP addresses
To find out whether an IP address already exists in the network, enter the
"PING" function in the input prompt (DOS window).
1. Switch on
components) which have been operated to date.
2. Invoke the input prompt (DOS window) from a network computer.
3. At the prompt, enter "ping" followed by the address to be checked and
then press the Enter key.
Example: "ping 192.168.15.13"
 If a network device responds, then this address has already been
assigned.
Installation Manual ORTHOPHOS XG 3D / Ceph
all
network devices (computers, printers, X-ray
Sirona Dental Systems GmbH
63 03 452 D3352
D3352.031.05.08.02 10.2015

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents