X-Ray Unit; System Safety - Sirona ORTHOPHOS XG 5 Maintenance Manual

Hide thumbs Also See for ORTHOPHOS XG 5:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
2.4

X-ray unit

X-ray tube assembly and stand condition
Date
(please enter month/year)
Tube assembly housing palpably/
visibly oil-free (Fig. 1)?
2.5

System safety

Safety
Date
(please enter month/year)
Coiled cable with strain relief and
anti-kink sleeve OK (Fig. 1)?
Power cable and 2nd ground wire
free of damage (Fig. 1)?
Control cable free of damage
(Fig. 1)?
All screws that fasten the X-ray
unit to the wall and the support
bracket tightly secured and/or unit
securely fastened to the stand
(Fig. 2)?
For Japan only: Is the unit imme-
diately switched off when the
emergency shutdown switch is
activated?
Anti-kink
sleeve
Fig. 1:
59 87 685 D 3352
D 3352.105.01.11.02
2 Annual inspection performed by the system owner or other authorized persons
If no  Inform your
Fig. 1:
If no  Immediately order repair by a service engineer.
Control cable
Power cable
2nd ground wire
Answer questions with yes () or no (−)
dealer
.
Tube assembly housing
Answer questions with yes () or no (−)
Fig. 2:
9