Amalgam-Test; Purge Function Purging The Water Paths - Sirona S2 Operating Instructions Manual

Table of Contents

Advertisement

Available languages

Available languages

Sirona Dental Systems GmbH
A
READY
10:04:27
A
SELECT MODE
PURGE / SANITATION
41 63 189 D 3311
D 3311.201.01.10.02
AMALG
A
SELECT ITEMS
FOR PURGING
A
PURGING
11 Unit lamp, Tumbler Filling, Cuspidor Rinsing, Purge
11.4

Amalgam-Test

If the tumbler filler and cuspidor rinsing
keys are briefly pressed simultaneously, an
acoustic signal sounds on the water unit
and AMALG flashes on the display.
Briefly pressing this key combination again
ends the test.
11.5
Purge Function
Purging the water paths
(only for dentist unit)
It is possible to flush through the water
paths with the Purge function (e.g. after a
lengthy break or after surgical treatments).
For this purpose press the two upper keys
in the left block, Cooling spray and Chip
blower..
A message will appear prompting you to
choose between PURGE and SANITA-
TION*. Select PURGE with the – or + key.
PURGE must flash.
* The SANITATION function is described in
the Maintenance Instructions.
Press the Chip blower key while PURGE
is still flashing. The message SELECT
ITEMS FOR PURGING then appears.
Set the sanitation cover on the cuspidor.
The water paths are purged basically
without the treatment instruments being
connected!
Pull the treatment instruments off from the
couplings.
Plug the couplings into the associated
adapters of the sanitation cover of the
same color.
Activate the Chip blower key anew, the
instrument hoses are then purged
consecutively.
The purging time set previously in Mode
runs for each instrument hose individually
(see page 35).
An acoustic signal sounds after the last
hose is purged.
Finally reconnect the instruments and
place them in their holders.