7.6

Removing water from the unit

You must remove the water from the unit if you will not be using it for a longer
period of time or wish to transport it.
No milling/scanning process is running.
1.
Open the flap on the front panel of the unit.
To open the front flap, pull it on both sides.
2.
Press the tank catch upward and carefully pull out the water tank toward
the front of the unit.
3.
Open the water tank, drain it and reinsert it.
4.
You can start the service program after installing the CEREC 3D program
in the "CEREC" program group. To do this, click "Start" / "Programs"
/ "CEREC" / "Service".
5.
You can also change the water of the milling unit without a service
password by clicking the "Maintenance only"button.
6.
Let the water pump continue running until no more water comes out of the
nozzles (approx. 1min.).
7.
Pull out the water tank and empty it.
8.
Push the water tank back into the housing just far enough so that the tank
catch engages (press the catch downward if necessary).