Removing Water From The Unit; Using The Tank Cap Opener - Sirona CEREC MC XL Operating Instructions Manual

Dental milling machine
Table of Contents

Advertisement

7 Maintenance

Removing water from the unit

46
NOTICE
Cleaning the filter
Clean the filter approx. every 12 to 15 units under running water, but at
least with every water change.
7.7
Removing water from the unit
You must remove the water from the unit if you will not be using it for a longer
period of time or wish to transpot it.
✔ No milling/scanning process is running.
1.
Switch the unit off.
2.
Pull out the water tank at the front of the unit.
3.
Drain the water out of the water tank through the drain opening and
reinsert the water tank in the unit.
4.
Switch the unit on.
NOTICE
Pump button active at power-up
"Pump"
The
button appears on the display when the milling unit is switched
on. You can start or stop the water pump by pressing this button.
"Pump"
5.
Press the
key to switch the pump on.
The water pump then starts pumping the water out of the unit.
Let the pump run until no more water escapes from the nozzles.
6.
Press the
"Pump"
key to switch the pump off.
7.
Pull out the water tank and empty it.
8.
Push it back into the housing.
7.8

Using the tank cap opener

CAUTION
Risk of damage to the tank
Use the tank cap opener only for opening the tank cap, tank drain and filter
insert.
Do not use the tank cap opener for closing the tank cap. To close the tank
cap, tank drain and filter insert it is sufficient to tighten them clockwise by
hand.
Sirona Dental Systems GmbH
Operating Instructions CEREC MC XL
61 46 885 D 3439
D 3439.201.02.11.02 10.2009