Removing Water From The Unit; Procedure For All Materials Except For Base Metals; Base Metal Operation Or Mixed Operation With Base Metal And Other; Sirona Dental Systems Gmbh - Sirona CEREC MC XL Operating Instructions Manual

Table of Contents

Advertisement

7 Service
7.7 Removing water from the unit
58
7.7

Removing water from the unit

7.7.1

Procedure for all materials except for base metals

You must remove water from the unit if you will not be using it for a long
period of time or wish to transport it.
No machining process is running.
1. Switch the unit off.
2. Pull out the water tank at the front of the unit.
3. Drain the water out of the water tank through the drain opening and
reinsert the water tank in the unit.
4. Switch the unit on.
NOTICE
Pump button active at power-up
"Pump"
The
button appears on the display when the grinding unit is
switched on. You can start or stop the water pump by pressing this
button.
"Pump"
5. Press the
 The water pump then starts pumping the water out of the unit.
Let the pump run until no more water escapes from the nozzles.
6. Press the
"Pump"
7. Pull out the water tank and empty it.
8. Push it back into the housing.
7.7.2
Base metal operation or mixed operation with base metal
and other materials
NOTICE
Fragment: Material manufacturer's safety information
Observe the safety information from the material manufacturer
Observe the safety instructions regarding occupational safety and
disposal referred to in the material manufacturer's operating
instructions. Dirty filters must be disposed of appropriately in
accordance with these specifications.
1. Empty the water tank (see Emptying the water tank [ → 45]) and insert
NEM inLab MC XL fragment
it back into the unit.
2. Switch the unit on.
"Pump"
3. Press the
 The water pump then starts pumping the water out of the unit.
Let the pump run until no more water escapes from the nozzles.
"Pump"
4. Press the
5. Pull out the water tank and empty it.
6. Push it back into the housing.

Sirona Dental Systems GmbH

Operating Instructions CEREC MC XL
key to switch the pump on.
key to switch the pump off.
button to switch the pump on.
button to switch the pump off.
D 3439.201.02.19.02 12.2014
61 46 885 D 3439

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents