:MTESt:AMASk:CREate
Command Syntax
:MTESt:AMASk:CREate
The :MTESt:AMASk:CREate command automatically constructs a mask
around the current selected channel, using the tolerance parameters
defined by the :MTESt:AMASk:XDELta, :MTESt:AMASk:YDELta, and
:MTESt:AMASk:UNITs commands. The mask only encompasses the portion
of the waveform visible on the display, so you must ensure that the
waveform is acquired and displayed consistently to obtain repeatable
results.
The :MTESt:SOURce command selects the channel and should be set before
using this command.
See Also
•
•
•
•
•
•
Example Code
•
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1138)
"Introduction to :MTESt Commands"
":MTESt:AMASk:XDELta"
":MTESt:AMASk:YDELta"
":MTESt:AMASk:UNITs"
":MTESt:AMASk:SOURce"
":MTESt:SOURce"
on page 521
"Example Code"
on page 491
on page 491
on page 498
on page 499
on page 497
on page 496
22
:MTESt Commands
495